Journal: EJNMMI Physics
Article Title: Harmonisation of standard uptake values in long-axial field-of-view PET/CT: an exploratory study
doi: 10.1186/s40658-025-00819-3
Figure Lengend Snippet: Proposed detector-splitting phantom scanning and harmonisation strategy for LAFOV PET. a Differences in scanning protocols and clinical workflows between LAFOV PET and SAFOV PET; b Scout image-based detector-splitting strategy. The long-axial detector is segmented into five positions (head, chest, abdomen, pelvis, and leg); c Phantom scanning performed sequentially at the different detector positions
Article Snippet: At baseline, a SAFOV PET/CT scan (Gemini 64, Philips, 18 F-FDG) in March 2025 demonstrated diffuse thickening of the gastric wall involving the gastric angle and antrum, with an SUV max of 21.30.
Techniques: